NSLS2/lsdc

NYX - sample centering - units

Closed this issue · 2 comments

  • click to center seems to be using wrong units for wrong situations - fix
  • focus changes appear to be very large - 5 mm. probably due to units mismatch
  • checkC2C_X also seems to be related to this issue - we were getting Click to center out of bounds on X move. Please mount next sample. when trying to center the sample.
  • possible solution: gonioUnits - mm or microns, then make sure all calculations are correct
    • needs to be noted in documentation

Focus change is not essential - should be handled by working on- and off-axis camera centering

However, click-to-center had issues when checkC2C_X failed, which made it impossible to center the sample, thus making the label "high"

working for both mags including click to center and focus buttons. nice work! closing!