NYX - sample centering - units
Closed this issue · 2 comments
JunAishima commented
- click to center seems to be using wrong units for wrong situations - fix
- focus changes appear to be very large - 5 mm. probably due to units mismatch
- checkC2C_X also seems to be related to this issue - we were getting
Click to center out of bounds on X move. Please mount next sample.
when trying to center the sample. - possible solution: gonioUnits - mm or microns, then make sure all calculations are correct
- needs to be noted in documentation
JunAishima commented
Focus change is not essential - should be handled by working on- and off-axis camera centering
However, click-to-center had issues when checkC2C_X failed, which made it impossible to center the sample, thus making the label "high"
JunAishima commented
working for both mags including click to center and focus buttons. nice work! closing!