martinloren/HScope

BUG: HS102 Pro Channel 2 AC/DC selection on reconnect

gadjex opened this issue · 5 comments

On an HS102 Pro, after turning on channel 2 to DC mode and then disconnecting the USB and reconnecting, the app will erroneously show DC mode on channel 2 when it is actually in AC mode. Switching channel 2 off and then back to DC mode will fix the issue until the next disconnect and
reconnect.

This is confirmed by using a volt meter and checking pin PB13 is at 0 volts after reconnecting.

Channel 1 does not display this improper behavior.

Thanks for reporting this issue. What is the firmware version?

Confirmed, it is a bug on HScope, it will be fixed in next Beta release. By the way there is no documentation for the HS102 Pro but the firmware is working probably....did you build it? I mean with the PGA and all range of input voltages or just single voltage range?

This issue should have been solved in the latest HScope release.