Ear tube & Y-piece outer diameter?
abitofalchemy opened this issue · 1 comments
abitofalchemy commented
- The tube connecting the eartube to the Y-pieace is called for it to be 20cm long in the list of materials, but if I look at the sample in FB it looks shorter, is 20 cm is till the length to go with?
- I had a really hard time with the outer diameter of ear-tubes and Y-piece and the 4mm inner diameter/6mm OD tubing. It seems to me that this tubing should have a wider ID or the Y and ear-tubes be smaller in diameter. What do you think? Look at the connecting points on my current prototype
tareko commented
Hi @abitofalchemy,
Congratulations on making the stethoscope! Regarding your questions:
- Yes, 20cm is really the maximum. If you want to wear it around your neck, that may even end up being a bit short., since the decreased weight will not properly hold down the stethoscope and it will jut to the sides.
- I strongly suggest the 4/6 and that is what we have tested with and is validated. However, if you want 6/8, there should theoretically be no deterioration.