billkarsh/SpikeGLX

BIST Test error 14

Closed this issue · 3 comments

Hi Bill,

Bother you again.

I am Qifan from Shanghai. Thanks for your help on Probe Config Files. I have already got them now.

However, there is still an issue on BIST test. When I run HST test with the dongle, all tests are successful. But for the BIST test. There is an error 14 on Test Parallel Serial Bus.

Do you have any idea about this?


Test Base Station
result = 0 'SUCCESS'


Test Heartbeat
Probe: open
Starting test, allow ~5 seconds...
result = 0 'SUCCESS'


Test SerDes Link
Probe: open
Starting test, allow ~10 seconds...
Test result: serDes error count = 0


Test I2C/MemMap
Probe: open
Starting test...
result = 0 'SUCCESS'


Test EEPROM
Probe: open
Starting test...
result = 0 'SUCCESS'


Test Shift Registers
Probe: open
Starting test...
result = 0 'SUCCESS'


Test Parallel Serial Bus
Probe: open
Starting test...
result = 14 'BIST_ERROR'


Test Signal Path
Probe: open
Starting test, allow ~40 seconds...
Signal test result = FAILED


Test Noise Level
Probe: open
Starting test, allow ~40 seconds...
result = 14 'BIST_ERROR'

Likely probe not seated or bad probe. Not a bug.

OK. Thanks, I will reconnect the probe and try again.