/simple_modularControl_v1.02

A streamlined fork of modularControl aimed at painless integration with new microscopes

Primary LanguageMATLAB

Simple modularControl


This a streamlined fork of the modularControl code. The goal here is to get your microscope up and running with minimum effort. New instruments can be easily configured and the GUI can be reconfigured as required. By design most of the back-end code is hidden from the end user, yet all the functionality of modularControl is retained.

Quick Links


 

Package summary:

A summary of the structure of this package with brief folder and file descriptors

* mcScope.m		`This is the main function that runs the package`

* mcUserInput.m		`All the instruments and devices are defined here`

* @mcgScope
	*       mcgScope.m			Wrapper __[Do not modify!]__
	*       ScopeConfig.m			The GUI elements are defined here (Galvo scan, Piezo scan, optimize, etc.)
	*       Callbacks.m			Define what action the GUI elements perform
	*       setupObjects.m			Ensure the GUI can interface with the instruments

* mcInstruments
	* @mcaDAQ	DAQ output devices	
		*	mcaDAQ.m		Wrapper __[Do not modify!]__
		*       piezoConfig.m		Piezo driver
		*       galvoConfig.m		Galvo driver
		*       analogConfig.m 		[template]
		*       digitalConfig.m		[template]	
	
	* @mciDAQ	DAQ input devices
		*       mciDAQ.m		Wrapper __[Do not modify!]__
		*       counterConfig.m		Counter input driver
		*       digitalConfig.m		[template]
		*       voltageConfig.m		[template]
	
	* @mcaMicro				Newport micrometer
		*       mcaMicro.m		Wrapper __[Do not modify!]__
		*       microConfig.m		Micrometer driver	
		
	* @mciNIGPIB				GPIB power meter 	[unused device]
		*       mciNIGPIB.m		Wrapper 
		*       powermeterConfig.m	power meter driver
	
	* @mciSpectrum				Python spectrometer 	[unused device]
		*       mciSpectrum.m		Wrapper 
		*       pyWinSpecConfig.m	spectrometer driver
		
	* @mciPFlip				Virtual instrument	[unused device]
		*       mciPFlip.m		Virtual Wrapper 
		*       pflipConfig.m		Virtual driver
		
	* extra		other unused device drivers and templates


* configs		During runtime, a GUI configuration is created and stored in this folder
* core			__[Do not modify!]__
* mcJoystick		Joystick functionality
* mcVideo		Interface to microscope live camera
* utility scripts 	Additional scripts that are used by the package such as peakfinder, etc.

Back to the top


 

How it works:

Here is how the default configuration can be described in blocks

alt text

  • The instrument driver defines how the physical instrument data (eg. in V) can be converted to digital data and vice versa.
  • The instrument wrapper defines how the digital data is converted to a containerized mc_object and vice versa.
    • For instruments that need to talk via the DAQ, the daq wrapper also takes care of multiple device synchronization (eg. synchronize counter, voltage output to piezos for a 2-D confocal scan)
  • The mc_object contains data and all the information required for user interactivity through the GUI (eg. type of input; say x-axis, integration time, etc.)

 

Here is how the GUI can be described in blocks

  • When the GUI is started, setupObjects.m is exectued and all the mc_objects are initialized (for all connected devices). Communication channels are established and the instruments are set to their default state (defined in their respective drivers).
  • Next, the user facing UI elements defined in ScopeConfig.m are executed and the GUI goes into standby-mode awaiting a user interaction event (i.e. a GUI callback).
  • When a UI element is triggered either by a mouse click or keypress event, the callback function for that specific UI elemnt (defined in Callbacks.m) is executed. This callback function can be a predefined function such as confocal scan or your very own custom function (ensure your function is in the matlab path).

Back to the top


 

How to configure your new microscope:

Download the package (simple_modularControl_vX.XX.zip) and extract the contents. Rename the extracted folder as 'simpleModularControl'. To run the package use mcScope.m.

The first time you run the package, you will have to click through three setup popups:

  1. Confirm the location of simpleModularControl folder. The package will try to determine its current location automatically. You can confirm if correct or pick a different folder if it is incorrect. Note: this is used to figure out the path to important core files, the package will fail to load if the wrong folder is selected.

  2. Confirm the loaction for background data saving. The package will automatically save all data to this background folder with the timestamp as the filename.

  3. Confirm the loaction for manual data saving. This is the folder that will open by default when you are saving manually.

Note: After completion of the initial setup, the selected options will be saved in the configs\your pc name\ folder as a .m file. If you see the setup popup during the second run then someting has gone wrong double check the folder selection.

 

To build a custom GUI for your microscope, follow these steps in order:

 

1. Write a driver for your device

A driver function defines the translation of commands from the GUI into a form that can be interpreted by your device. You can think of this as the physcial hardware abstraction layer of the package.

  • Figure out what devices are connected to your microscope (eg. Piezos, Galvos, spectrometer, etc.).
  • Understand how to talk to these devices. Always check the manual or refer to the example code. Matlab instrument control app or NI-Max is a good place to test instrument control.
  • Encapsulate device communication protocol into a driver function (there are templates under mcInstruments->extras).

The wrapper function will use the hardware protocol defined in the driver to communicate with your device. This makes it simple to reconfigure the GUI in case the devices are disconnected and reconnnected to a different hardware port.

See below for a outline of a driver with a brief description of the parameters (this example uses piezoConfig.m which is a driver for piezo control through the DAQ input/output instrument):  

%Piezo-systems Jena Nanopositioner
%Currently installed on the NV microscope

function config = piezoConfig()
    config.class =              'mcaDAQ';

    config.name =               'Default Piezo';

    config.kind.kind =          'NIDAQanalog'; 		  % Actual hardware interface
    config.kind.name =          'PiezosystemsJena Piezo'; % For user info. only
    config.kind.intRange =      [0 10];			  % Output voltage range
    config.kind.int2extConv =   @(x)(8.*(5 - x));         % Conversion from 'internal' units to 'external'.
    config.kind.ext2intConv =   @(x)((40 - x)./8);        % Conversion from 'external' units to 'internal'.
    config.kind.intUnits =      'V';                      % 'Internal' units.
    config.kind.extUnits =      'um';                     % 'External' units.
    config.kind.base =           5;                       % The (internal) value that the axis seeks at startup.

    config.dev =                'Dev1';			  % DAQ hardware address
    config.chn =                'ao0';			  % DAQ analog output channel
    config.type =               'Voltage';		  % DAQ output type 

    config.keyStep =            .1;			  % Default step size 
    config.joyStep =            .5;			  % Default step size (if joystick exists) 	
end

 

More details on parameters defined within the included drivers:

Show all details

For your reference (it is always a good ideal to keep your code readable!)
config.kind.kind the programatic name of the device (i.e. type of interface, physical device identifier, etc.)
config.kind.name the explanatory name of the device (i.e. manufacturer, model number, etc.)

Common Parameters
These paramters are device agnostic

  • In a mcAxis type driver [i.e. for bi-directional communication; Input and Output thru DAQ]
    config.kind.intUnits a string representing the units that the axis uses internally (e.g. for piezos, this is volts)
    config.kind.extUnits a string representing the units that the user should use (e.g. for piezos, this is microns)
    config.kind.int2extConv conversion from internal to external units (e.g. for piezos, 0V maps to -25um, 10V maps to 25um)
    config.kind.ext2intConv conversion from external to internal units, this should be the inverse of int2extConv
    config.kind.intRange the range of the axis in external units (this is generated using the conversions)
    config.kind.extRange the range of the axis in external units (this is generated from intRange using the conversions)
    config.kind.base the value (in internal units) that the axis should seek at startup

  • In a mcInput type driver [i.e. for Input only; eg. counter input through DAQ]
    config.kind.extUnits the appropriate units (no internal units are neccessary here)
    config.kind.shouldNormalize whether or not the measurement should be divided by the time taken to measure
    config.kind.sizeInput the expected size of the input (this allows other parts of the program to allocate space for the mcInput before the measurement has been taken for numbers, this is set to [1 1]; for a vector like a spectrum, this could be [512 1]).

Device specific Parameters
These parameters are utilized by your wrapper to figure out the hardware communication channel

  • For a DAQ device
    config.dev a string that identifies the DAQ. Check NI-MAX for the identifier (deafult is 'Dev1' can vary if you have usbDAQ or multiple DAQs)
    config.chn a string that identifies the DAQ channel (eg. for analog output on daq channel-0 use 'ao0')
    config.type a string defining the type of DAQ channel (eg. for voltage ouput use 'Voltage')

  • For a USB device (eg. newport micrometer)
    config.port a string that identifies the USB serial port for the micrometer controller (eg.'COM4')
    config.addr a string required by the newport usb controller in case multiple micrometers are connected a single usb controller (default '1')

  • For a custom device
    You may have to define new parmaters for your device.

Note: The physical hardware uses internal units whereas the user uses external units. For instance, a piezo uses Volts internally but microns externally. The external units are defined via the anonymous function config.kind.int2extConv and its inverse config.kind.ext2intConv. For instance, for the piezos that we use in the diamond room, we convert between a 0 to 10 V range and a -25 to 25 um range. Thus,

config.kind.int2extConv = @(x)(5.*(5 - x));
config.kind.ext2intConv = @(x)((25 - x)./5);

Keep in mind that the internal mcObject variables a.x and a.xt --- the current and target positions --- use internal units. The external current and target positions can be found via a.getX() and a.getXt().

Back to the top

 

2. Write a wrapper for your driver

A wrapper is a class that utilizes the driver function to perform all communication with the device and converts runtime data into an mc_object container. The driver functions from step#1 are inherited as a static method within the wrapper class definition.

Note: If you are using a pre-existing wrapper, please ensure that your new driver is registered as a static method within the wrapper.

 

See below for a outline of a wrapper with a brief description of the methods (this example uses mcaDAQ.m which is a wrapper for the DAQ input/output instrument):

classdef mcaDAQ < mcAxis 
% Make sure that the folder name is @class_name. This folder should only contain the specific instrument driver + wrapper
% There are additional methods inherited from mcAxis that generally do not need to be modified by the user 

    methods (Static)    
   	% Here you need to register your driver functions defined in step #1
   	% There are two devices connected to the DAQ
        mc_object = piezoConfig(); %<- register the piezo driver with the wrapper
        mc_object = galvoConfig(); %<- register the galvo driver with the wrapper     
    
    methods ()  
        function Open(mc_object)  	%<- Open a communication channel with the DAQ.	
        function Close(mc_object)	%<- Close the communication channel with the DAQ.
        function Goto(mc_object, x)     %<- Communicate with the device; send x

        % EXTRA
	% This is only required for DAQ wrappers
	% required to synchronize data acquisition between multiple devices connected to the DAQ   
        function addToSession(mc_object, s)	 
end

Back to the top

 

3. Virtual Instruments

Some experiments require orchestration of multiple devices. In such cases you can define virtual drivers and wrappers that can encapsulate an instance of the real devices, with the experiment details being stored in a single mc_object. For example, if you need to actuate a servo controlled flip-mirror communicating via the DAQ digitalOutput and simultaneously readout the laser power from a GPIB power meter device. The individual drivers and wrappers for the flip-mirror and powermeter can be combined into a mciPFlip virtual instrument.

 

Such an example is included in mcInstruments and is outlined below:

Show all details

Outline of the pflipconfig virtual driver:

function config = pflipConfig()
	config.class = 'mciPFlip';	
	config.name = 'Powermeter';
	
	% using an instance of the predefined wrapper + driver for NI-GPIB instrument
	config.power		= 	mciNIGPIB(); 
	
	% using an instance of the predefined wrapper + driver for NI-DAQ instrument
	flipConfig 		=       mcaDAQ.digitalConfig();
	flipConfig.chn 		= 	'Port0/Line1'; 
	flipConfig.name 	= 	'Flip Mirror'; 	
	config.flip		=       mcaDAQ(flipConfig);

	config.kind.kind    	=       'mciPFlip';
	config.kind.name    	=       'Interjecting Powermeter';
	
	% you can inherit properties from other instruments
	config.kind.extUnits 	=       config.power.config.kind.extUnits;   
end

Outline of the mcPFlip virtual wrapper:

classdef mciPFlip < mcInput
    methods (Static)         
        config = pflipConfig();		
    end
    
    methods
        % OPEN/CLOSE
        function Open(mc_object)
            mc_object.config.power.open();
            mc_object.config.flip.open();
        end
	
        function Close(mc_object)
            mc_object.config.power.close();
            mc_object.config.flip.close();
        end
        
        % MEASURE
        function data = Measure(mc_object, ~)
            prevX = mc_object.config.flip.getX();       % Get the previous state of the flip mirror.
            
            mc_object.config.flip.goto(0);      	% Flip the mirror in...
            pause(.5);                         		% Wait a bit...
            data = mc_object.config.power.measure();    % And measure.
            
            mc_object.config.flip.goto(prevX);          % Then restore the previous state.
        end
end

 

Note: You can perform the operation described above with a simple custom function. However a virtual instrument can be much more complex. Look at mcInstruments->@mciPLE that combines a DAQ counter input, DAQ analog out (voltage for laser wavelength control), 2x DAQ digital out (laser shutter and servo ND filters). With the virtual instrument, additional methods can be integrated for plotting, processing and analyzing the acquired PLE data.

 

Keep in mind that there are no limitiation on creating new virtual instruments from combination of other virtual instruments! Additionally, in case of any changes to the physical hardware connections, all that needs to be updated are the protocol definitions in the base driver files :)

Back to the top

 

4. Initialization

There are two files that have to be modified for the initialization process:

  1. Configure all the devices in mcUserInput.m. This file defines all the manual instrument controls.
    After defining all the instruments and device variants (i.e. Piezo-X, Piezo-Y, etc. that use the same driver and wrapper but different communication channel/hardware port), add the instruments to the config.axesGroups cell array. Please follow the detailed comments in the file to reconfigure for your microscope. The manual user control GUI panel can be configured in the function makeGUI(varin) method.

  2. Configure device for automated tasks such as confocal xy scanning, PLE, ODMR, etc in @mcgScope->setupObjects.m. Here we can utilize all the instruments and device variants defined in the previous step. We import the 'config.axesGroups' from mcUserInput.m and assign the instruments and devices to the mc_obj instances.

    Note: Keep in mind that if you modify config.axesGroups in mcUserInput.m ensure that @mcgScope->setupObjects.m is modified accordingly. Please pay close attention to the cell indexing. If the device configurations in the two files do not match, you will recieve an error when running any automated task (like confocal).

Back to the top

 

5. User interface

WIP

Add instrument elements to the UI in ScopeConfig.m.
Define the actions performed by the new UI elements in Callbacks.m