This a streamlined fork of the modularControl code. The goal here is to get your microscope up and running with minimum effort. New instruments can be easily configured and the GUI can be reconfigured as required. By design most of the back-end code is hidden from the end user, yet all the functionality of modularControl is retained.
A summary of the structure of this package with brief folder and file descriptors
* mcScope.m `This is the main function that runs the package` * mcUserInput.m `All the instruments and devices are defined here` * @mcgScope * mcgScope.m Wrapper __[Do not modify!]__ * ScopeConfig.m The GUI elements are defined here (Galvo scan, Piezo scan, optimize, etc.) * Callbacks.m Define what action the GUI elements perform * setupObjects.m Ensure the GUI can interface with the instruments * mcInstruments * @mcaDAQ DAQ output devices * mcaDAQ.m Wrapper __[Do not modify!]__ * piezoConfig.m Piezo driver * galvoConfig.m Galvo driver * analogConfig.m [template] * digitalConfig.m [template] * @mciDAQ DAQ input devices * mciDAQ.m Wrapper __[Do not modify!]__ * counterConfig.m Counter input driver * digitalConfig.m [template] * voltageConfig.m [template] * @mcaMicro Newport micrometer * mcaMicro.m Wrapper __[Do not modify!]__ * microConfig.m Micrometer driver * @mciNIGPIB GPIB power meter [unused device] * mciNIGPIB.m Wrapper * powermeterConfig.m power meter driver * @mciSpectrum Python spectrometer [unused device] * mciSpectrum.m Wrapper * pyWinSpecConfig.m spectrometer driver * @mciPFlip Virtual instrument [unused device] * mciPFlip.m Virtual Wrapper * pflipConfig.m Virtual driver * extra other unused device drivers and templates * configs During runtime, a GUI configuration is created and stored in this folder * core __[Do not modify!]__ * mcJoystick Joystick functionality * mcVideo Interface to microscope live camera * utility scripts Additional scripts that are used by the package such as peakfinder, etc.
Here is how the default configuration can be described in blocks
- The instrument
driverdefines how the physical instrument data (eg. in V) can be converted to digital data and vice versa. - The instrument
wrapperdefines how the digital data is converted to a containerizedmc_objectand vice versa.- For instruments that need to talk via the DAQ, the daq wrapper also takes care of multiple device synchronization (eg. synchronize counter, voltage output to piezos for a 2-D confocal scan)
- The
mc_objectcontains data and all the information required for user interactivity through the GUI (eg. type of input; say x-axis, integration time, etc.)
Here is how the GUI can be described in blocks
- When the GUI is started,
setupObjects.mis exectued and all themc_objectsare initialized (for all connected devices). Communication channels are established and the instruments are set to their default state (defined in their respective drivers). - Next, the user facing UI elements defined in
ScopeConfig.mare executed and the GUI goes into standby-mode awaiting a user interaction event (i.e. a GUI callback). - When a UI element is triggered either by a mouse click or keypress event, the callback function for that specific UI elemnt (defined in
Callbacks.m) is executed. This callback function can be a predefined function such as confocal scan or your very own custom function (ensure your function is in the matlab path).
Download the package (simple_modularControl_vX.XX.zip) and extract the contents. Rename the extracted folder as 'simpleModularControl'. To run the package use mcScope.m.
The first time you run the package, you will have to click through three setup popups:
-
Confirm the location of simpleModularControl folder. The package will try to determine its current location automatically. You can confirm if correct or pick a different folder if it is incorrect. Note: this is used to figure out the path to important core files, the package will fail to load if the wrong folder is selected.
-
Confirm the loaction for background data saving. The package will automatically save all data to this background folder with the timestamp as the filename.
-
Confirm the loaction for manual data saving. This is the folder that will open by default when you are saving manually.
Note: After completion of the initial setup, the selected options will be saved in the
configs\your pc name\folder as a .m file. If you see the setup popup during the second run then someting has gone wrong double check the folder selection.
To build a custom GUI for your microscope, follow these steps in order:
A driver function defines the translation of commands from the GUI into a form that can be interpreted by your device. You can think of this as the physcial hardware abstraction layer of the package.
- Figure out what devices are connected to your microscope (eg. Piezos, Galvos, spectrometer, etc.).
- Understand how to talk to these devices. Always check the manual or refer to the example code. Matlab instrument control app or NI-Max is a good place to test instrument control.
- Encapsulate device communication protocol into a driver function (there are templates under
mcInstruments->extras).
The wrapper function will use the hardware protocol defined in the driver to communicate with your device. This makes it simple to reconfigure the GUI in case the devices are disconnected and reconnnected to a different hardware port.
See below for a outline of a driver with a brief description of the parameters (this example uses piezoConfig.m which is a driver for piezo control through the DAQ input/output instrument):
%Piezo-systems Jena Nanopositioner
%Currently installed on the NV microscope
function config = piezoConfig()
config.class = 'mcaDAQ';
config.name = 'Default Piezo';
config.kind.kind = 'NIDAQanalog'; % Actual hardware interface
config.kind.name = 'PiezosystemsJena Piezo'; % For user info. only
config.kind.intRange = [0 10]; % Output voltage range
config.kind.int2extConv = @(x)(8.*(5 - x)); % Conversion from 'internal' units to 'external'.
config.kind.ext2intConv = @(x)((40 - x)./8); % Conversion from 'external' units to 'internal'.
config.kind.intUnits = 'V'; % 'Internal' units.
config.kind.extUnits = 'um'; % 'External' units.
config.kind.base = 5; % The (internal) value that the axis seeks at startup.
config.dev = 'Dev1'; % DAQ hardware address
config.chn = 'ao0'; % DAQ analog output channel
config.type = 'Voltage'; % DAQ output type
config.keyStep = .1; % Default step size
config.joyStep = .5; % Default step size (if joystick exists)
end
More details on parameters defined within the included drivers:
Show all details
For your reference (it is always a good ideal to keep your code readable!)
config.kind.kind the programatic name of the device (i.e. type of interface, physical device identifier, etc.)
config.kind.name the explanatory name of the device (i.e. manufacturer, model number, etc.)
Common Parameters
These paramters are device agnostic
-
In a
mcAxistype driver [i.e. for bi-directional communication; Input and Output thru DAQ]
config.kind.intUnitsa string representing the units that the axis uses internally (e.g. for piezos, this is volts)
config.kind.extUnitsa string representing the units that the user should use (e.g. for piezos, this is microns)
config.kind.int2extConvconversion from internal to external units (e.g. for piezos, 0V maps to -25um, 10V maps to 25um)
config.kind.ext2intConvconversion from external to internal units, this should be the inverse ofint2extConv
config.kind.intRangethe range of the axis in external units (this is generated using the conversions)
config.kind.extRangethe range of the axis in external units (this is generated fromintRangeusing the conversions)
config.kind.basethe value (in internal units) that the axis should seek at startup -
In a
mcInputtype driver [i.e. for Input only; eg. counter input through DAQ]
config.kind.extUnitsthe appropriate units (no internal units are neccessary here)
config.kind.shouldNormalizewhether or not the measurement should be divided by the time taken to measure
config.kind.sizeInputthe expected size of the input (this allows other parts of the program to allocate space for themcInputbefore the measurement has been taken for numbers, this is set to[1 1]; for a vector like a spectrum, this could be [512 1]).
Device specific Parameters
These parameters are utilized by your wrapper to figure out the hardware communication channel
-
For a DAQ device
config.deva string that identifies the DAQ. Check NI-MAX for the identifier (deafult is 'Dev1' can vary if you have usbDAQ or multiple DAQs)
config.chna string that identifies the DAQ channel (eg. for analog output on daq channel-0 use 'ao0')
config.typea string defining the type of DAQ channel (eg. for voltage ouput use 'Voltage') -
For a USB device (eg. newport micrometer)
config.porta string that identifies the USB serial port for the micrometer controller (eg.'COM4')
config.addra string required by the newport usb controller in case multiple micrometers are connected a single usb controller (default '1') -
For a custom device
You may have to define new parmaters for your device.
Note: The physical hardware uses internal units whereas the user uses external units. For instance, a piezo uses Volts internally but microns externally. The external units are defined via the anonymous function
config.kind.int2extConvand its inverseconfig.kind.ext2intConv. For instance, for the piezos that we use in the diamond room, we convert between a 0 to 10 V range and a -25 to 25 um range. Thus,config.kind.int2extConv = @(x)(5.*(5 - x));
config.kind.ext2intConv = @(x)((25 - x)./5);Keep in mind that the internal
mcObjectvariablesa.xanda.xt--- the current and target positions --- use internal units. The external current and target positions can be found viaa.getX()anda.getXt().
A wrapper is a class that utilizes the driver function to perform all communication with the device and converts runtime data into an mc_object container. The driver functions from step#1 are inherited as a static method within the wrapper class definition.
Note: If you are using a pre-existing wrapper, please ensure that your new driver is registered as a static method within the wrapper.
See below for a outline of a wrapper with a brief description of the methods (this example uses mcaDAQ.m which is a wrapper for the DAQ input/output instrument):
classdef mcaDAQ < mcAxis
% Make sure that the folder name is @class_name. This folder should only contain the specific instrument driver + wrapper
% There are additional methods inherited from mcAxis that generally do not need to be modified by the user
methods (Static)
% Here you need to register your driver functions defined in step #1
% There are two devices connected to the DAQ
mc_object = piezoConfig(); %<- register the piezo driver with the wrapper
mc_object = galvoConfig(); %<- register the galvo driver with the wrapper
methods ()
function Open(mc_object) %<- Open a communication channel with the DAQ.
function Close(mc_object) %<- Close the communication channel with the DAQ.
function Goto(mc_object, x) %<- Communicate with the device; send x
% EXTRA
% This is only required for DAQ wrappers
% required to synchronize data acquisition between multiple devices connected to the DAQ
function addToSession(mc_object, s)
end
Some experiments require orchestration of multiple devices. In such cases you can define virtual drivers and wrappers that can encapsulate an instance of the real devices, with the experiment details being stored in a single mc_object. For example, if you need to actuate a servo controlled flip-mirror communicating via the DAQ digitalOutput and simultaneously readout the laser power from a GPIB power meter device. The individual drivers and wrappers for the flip-mirror and powermeter can be combined into a mciPFlip virtual instrument.
Such an example is included in mcInstruments and is outlined below:
Show all details
Outline of the pflipconfig virtual driver:
function config = pflipConfig()
config.class = 'mciPFlip';
config.name = 'Powermeter';
% using an instance of the predefined wrapper + driver for NI-GPIB instrument
config.power = mciNIGPIB();
% using an instance of the predefined wrapper + driver for NI-DAQ instrument
flipConfig = mcaDAQ.digitalConfig();
flipConfig.chn = 'Port0/Line1';
flipConfig.name = 'Flip Mirror';
config.flip = mcaDAQ(flipConfig);
config.kind.kind = 'mciPFlip';
config.kind.name = 'Interjecting Powermeter';
% you can inherit properties from other instruments
config.kind.extUnits = config.power.config.kind.extUnits;
endOutline of the mcPFlip virtual wrapper:
classdef mciPFlip < mcInput
methods (Static)
config = pflipConfig();
end
methods
% OPEN/CLOSE
function Open(mc_object)
mc_object.config.power.open();
mc_object.config.flip.open();
end
function Close(mc_object)
mc_object.config.power.close();
mc_object.config.flip.close();
end
% MEASURE
function data = Measure(mc_object, ~)
prevX = mc_object.config.flip.getX(); % Get the previous state of the flip mirror.
mc_object.config.flip.goto(0); % Flip the mirror in...
pause(.5); % Wait a bit...
data = mc_object.config.power.measure(); % And measure.
mc_object.config.flip.goto(prevX); % Then restore the previous state.
end
end
Note: You can perform the operation described above with a simple custom function. However a virtual instrument can be much more complex. Look at
mcInstruments->@mciPLEthat combines a DAQ counter input, DAQ analog out (voltage for laser wavelength control), 2x DAQ digital out (laser shutter and servo ND filters). With the virtual instrument, additional methods can be integrated for plotting, processing and analyzing the acquired PLE data.
Keep in mind that there are no limitiation on creating new virtual instruments from combination of other virtual instruments! Additionally, in case of any changes to the physical hardware connections, all that needs to be updated are the protocol definitions in the base driver files :)
There are two files that have to be modified for the initialization process:
-
Configure all the devices in
mcUserInput.m. This file defines all the manual instrument controls.
After defining all the instruments and device variants (i.e. Piezo-X, Piezo-Y, etc. that use the same driver and wrapper but different communication channel/hardware port), add the instruments to theconfig.axesGroupscell array. Please follow the detailed comments in the file to reconfigure for your microscope. The manual user control GUI panel can be configured in thefunction makeGUI(varin)method. -
Configure device for automated tasks such as confocal xy scanning, PLE, ODMR, etc in
@mcgScope->setupObjects.m. Here we can utilize all the instruments and device variants defined in the previous step. We import the 'config.axesGroups' frommcUserInput.mand assign the instruments and devices to the mc_obj instances.Note: Keep in mind that if you modify
config.axesGroupsinmcUserInput.mensure that@mcgScope->setupObjects.mis modified accordingly. Please pay close attention to the cell indexing. If the device configurations in the two files do not match, you will recieve an error when running any automated task (like confocal).
WIP
Add instrument elements to the UI in ScopeConfig.m.
Define the actions performed by the new UI elements in Callbacks.m