Welcome to the OpenVentilator project. This is an OpenSource Ventilator / Mechanical Respirator for the Covid-19 Crisis.
This initiative as well as OpenLung was born on the Open Source COVID19 Medical Supplies Facebook Group by the awareness on creating a Ventilator solution for the scarcity plaguing our society world wide. I contacted Jeremias Almada from Argentina who by that time had presented an Ambu solution and a Cad Design. The idea was interseting but needed improvments
Since than we tryied to stablish some development and community standards and changed several times the project
PROJECT STATUS: We still need validation with health regulatory institutions and compliance with clinical requirements.
PLEASE DO NOT USE ONLY TIRE-CHAMBER, ENCASE THE the ineer part of the TireChamber with a foodbag like ZIPLOCK or any material with compliance to contact with food.
Design, Build, Validate and Supply a reliable Ventilation Medical Equipment for people/regions/countries in dificult economical situations with a component and mechanical agnostic philosophy. (That's is why we didn't continued putting efforts on the projects being developed by other teams who have a different society and economical reality)
The equipment must have as few industrial parts as possible. If necessary, industrial parts must be easily accessible, even in small towns and villages.
The equipment should be built independent of the main motor or the ventilation tool (AmbuBag, Bellow etc) to increase modularity of the parts and resources
The equipment can be built with as few tools as possible, and even with the use of scrap to facilitate access to materials.
Thechnical features are still being decided, the ventilator has not being validated yet.
Especification | Spartan model | Mark II |
---|---|---|
Volume Control | mechanical native feature | mechanical native feature |
Ventilation Frecuency | implementing 2 types | electronic native feature |
Inhale/Exhale proportion | implementing 2 types | electronic native feature |
Inlet air Filtration | Hepa(recomended) | Mandatory HEPA |
Inlet air humidifier | Aspirator(recomended) | Mandatory -Aspirator |
Exahaust filtration | Optional(recomended) | Mandatory |
Peep Pressure regulator | House Pipe solution | House pipe solution |
Peep Preassure indicator | Water Column indicator | Digital UI or WaterColumn |
Oxymeter | Not available | Aditional Hardware |
- PEEP Valve control between 1mm/h20 ~ 40mm/h20
- Mechanical Volumetric Ventilation Adjustment
- Contagion reduction by contaminated air
- Brazil Sanitary Regulations
Validated Prototype (Looking for partners for testing on Lung Simulators)
- Electronic Controller Status
Testing
- Mechanical Motor Status
Implementing
- Ventilator Status:
Validation
- Humidifier Status:
Design
- Filtering status
Implementing
- Peep Valve Status:
Adopted
- Thanks to ProjectOpenAir - Breathing tube Status:
Testing
For more information: https://www.popsolutions.co/openventilator
First, ** CLICK HERE ** to complete the form please, so we can organize everybody
Then join the whatsappgroup and talk with amanda (+55 11 99735-5042 ): https://chat.whatsapp.com/HRMx9xzVdt8Gpmwgm7ZVZ3
Documentation, Hardware, Design and Code discussions This was depracated to avoid outsiders looking for profit
Check the decissions ALREADY MADE
Speciall thanks to:
- Jeremias Almada
- Fabian Franz
- Jose Ignácio Méndez
- Washington Perez
- Jaqueline Passos
- Amanda Pellini
- Marguel Gutierrez
- Henrique Aguilar
- Vandeir Soares
- Ronaldo Alves
- Wendell Mendes
- Rodrigo Borges
- Henrique Nery
- Duit
- Carlos Delfino
- Marcio Dultra
- Três meninas hardware store
- The MIT guys from this paper: Abdul Mohsen Al Husseini, Heon Ju Lee, Justin Negrete, Stephen Powelson, Amelia Servil, Alexander Slocum, Jussi Saukkonen.
All our families, wifes and husbands that for the last days have been supporting us on our crazyness.
All the doctors, nurses and paramedics on the field fighting this common enemy