This is a collection of various Pulseq tutorials, which is supposed to grow in future

Primary LanguageMATLABCreative Commons Zero v1.0 UniversalCC0-1.0

Welcome to the Pulseq tutorials repository!

This is a collection of various Pulseq tutorials, which is supposed to grow in future. If you don't know where to start, read through this document and study the Pulseq Concepts presentation. Thereafter work through the tutorials one-by-one. Please let us know if your favorite topic is missing or needs more attention, we will do our best to cover it in future.

Foreword

The Pulseq1 software provides an open-source framework for the development, analysis, display, and execution of magnetic resonance (MR) sequences. To familiarise yourself with the Pulseq software, we recommend starting your Pulseq journey with this Pulseq Tutorial. It was initially developed for the Pulseq software demonstration and hands-on session at the Italian Chapter of ISMRM 2022 in Pisa and a few other live demos. This Tutorial contains example pulse sequences, raw MR data in the Siemens TWIX format, and reconstruction scripts, ranging from a free induction decay (FID) sequence to non-Cartesian radial and spiral sequences, in the order of increasing complexity. This document provides detailed instructions on how to work through the Tutorial. After going through this guided tour, you will probably appreciate both the simplicity and the power of Pulseq and be able to move on to building your own sequences with Pulseq!

If you come across any problems, please contact our Pulseq team: pulseq.mr-at-uniklinik-freiburg.de

Naming convention of the tutorials in the collection

The tutorials are ordered by increasing complexity, so if unsure where to start, 01_... is a good option.

In more detail, the naming convention is

LN_Topic

where

  • L is a one-digit level going from 0 for beginner to 1 for intermediate to 2 advanced;
  • N is a sequential single-digit number (this assumes we can only have up to 9 tutorials per complexity level);
  • Topic is a free text briefly describing the scope of the tutorial.

For the absolute beginners 01_from_FID_to_PRESS is a good start, which describes how to code sequences starting from a most basic FID pulse-acquire sequence moving on towards a reasonably optimised point-resolved spectroscopy (PRESS) sequence.

Author and version

This tutorial was written by Qingping Chen (with some edits by Maxim Zaitsev)

Division of Medical Physics, Department of Radiology, University Medical Center Freiburg, Germany

Initial revision created on June 21 2023

Contents

Foreword

1. Prerequisites

1.1 MR Physics Background

1.2 Programming Tool

1.3 Pulseq Software

1.4 mapVBVD Software

1.5 Text Compare Tool

2. Introduction to Pulseq

2.1 Matlab and Pulseq Sequence Files

2.2 Pulseq Interpreter

2.3 Blocks and Events in Pulseq

3. Tutorials Overview

From_FID_to_PRESS

Basic_gradient_echo

From_GRE_to_EPI

Radial_and_nonCartesian

1. Prerequisites

1.1 MR Physics Background

To be able to code MR sequences, knowledge in MR physics is required. We recommend some useful books and websites listed below.

Books:

Websites:

1.2 Programming Tools

1.3 Pulseq Software

Please download the Pulseq software from https://github.com/pulseq/pulseq. Install Pulseq software in Matlab by adding its directory and subdirectories to Matlab's path.

1.4 mapVBVD Software

The mapVBVD software is required to read raw data in Siemens TWIX format. The software can be downloaded from https://github.com/pehses/mapVBVD and installed by adding its directory to Matlab's path.

1.5 Text Compare Tool

We recommend using a text comparison tool to compare the sequences within the subsequent steps to visualise the changes that occur at each step. Meld software can be used for text comparison and can be downloaded from https://meldmerge.org/. By using Meld or other comparable software packages, it is possible to very quickly make the relevant changes easily visible, as in the example below.

Meld_Comparison

2. Introduction to Pulseq

Pulseq is an open-source framework for the development, analysis, display, and execution of MR sequences for imaging and spectroscopy. Under this framework, one can directly program an MR sequence in Matlab or Python and execute the Pulseq sequence file on real scanner hardware through the corresponding Pulseq interpreter. A central contribution of Pulseq is that it provides an open file format to compactly describe MR pulse sequences suitable for execution on a real MR scanner. The detailed file specification can be obtained from https://pulseq.github.io/specification.pdf.

2.1 Matlab and Pulseq Sequence Files

For the generation of Pulseq sequence files that are readable by an MR scanner, we run Matlab scripts (extension: .m), e.g. s01_FID.m. When a Matlab script runs successfully, a Pulseq sequence file is created (extension: .seq), e.g. s01_FID.seq. This file defines all sequence timing and events, including RF, delays, trapezoid gradients, arbitrary gradients, and ADC events. This forms a complete description of the sequence and is ideal for rapid prototyping and executing sequences programmed in alternative environments (e.g. JEMRIS). This text file can be read by the Pulseq interpreter on the scanner and stored as compressed sequence blocks in memory. Matlab and Python toolboxes also provide functionality to read Pulseq .seq files.

2.2 Pulseq Interpreter

The Pulseq interpreter can be used to execute the Pulseq sequence file (.seq file) generated by Matlab, PyPulseq or other tools on a particular MR scanner. The interpreter for Siemens scanners can be obtained after signing a Siemens C2P agreement (for more information, please send an inquiry to pulseq.mr-at-uniklinik-freiburg.de). The source code contained in the interpreter package can be compiled in the Siemen IDEA environment, and then the .so and .dll files are produced. Copy both binary files to the sequence folder on the MR scanner (folder defined by the %CustomerSeq% variable, C:\MedCom\MriCustomer\seq for Numaris4 C:\ProgramData\Siemens\Numaris\MriCustomer\CustomerSeq for NumarisX). These binary files of the Pulseq interpreter sequence are universal and can be used for any .seq file, so they only need to be copied to the scanner once. The .seq file that you want to execute needs to be placed in the subfolder %CustomerSeq%\pulseq. By default, the file named external.seq is used. Important: for the default protocol corresponding to the Pulseq interpreter sequence, a valid .seq file needs to be placed in %CustomerSeq%\pulseq stored under the name external.seq. To access the Pulseq interpreter sequence on the MR scanner, you need to create the corresponding protocol and eventually store it in your favourite examination folder. Navigate to Dot Cockpit->Program Editor->Browse->Default->Sequence Region->Customer sequences->Default. Now insert the Pulseq sequence into any protocol. Run this protocol and it will automatically read the Pulseq file external.seq in the folder %CustomerSeq%\pulseq. Refer to the Pulseq C2P documentation for further details.

As for the Pulseq interpreter for GE scanners, Please go to https://toppemri.github.io/ for more details.

2.3 Blocks and Events in Pulseq

Under the specification of the Pulseq file format, a sequence is established based on at least one block (id_block>0). Each entry in the block table declares a single block specified by a duration (DUR) in the units of BlockDurationRaster and a list of six event IDs:

<id_block> <dur> <rf_id> <gx_id> <gy_id> <gz_id> <adc_id> <ext_id>

The six events include an RF event (RF), three gradient events (GX, GY, GZ), an ADC event (ADC), and an extension event (EXT). An event ID of 0 indicates no event.

The RF events are declared in a section with the [RF] keyword. Each line in the section is specified by eight numbers:

<rf_id> <amp> <mag_id> <phase_id> <time_id> <delay> <freq_offset> <phase_offset>

The GRADIENT events are declared in two sections. Arbitrary gradients are located in the section marked with the [GRADIENTS] keyword. Each line in the section is specified by five numbers:

<grad_id> <amp> <shape_id> <time_id> <delay>

Trapezoidal gradients are located in the section labeled with the [TRAP] keyword. Each line in this section is specified by six numbers:

<grad_id> <amp> <rise> <flat> <fall> <delay>

As you may have noticed, arbitrary gradients and trapezoidal gradients share the same ID space - grad_id that uniquely identifies every gradient, with each ID only allowed to be present either in the [GRADIENTS] section or in the [TRAP] section.

The ADC section is declared with the [ADC] keyword. Each line in this section is specified by six numbers:

<adc_id> <num> <dwell> <delay> <freq_offset> <phase_offset>

The EXTENSION section is declared with the [EXTENSIONS] keyword. Each line in the section is specified by four numbers:

<ext_id> <type> <ref> <next>

Take one of the simplest MR experiments -- free induction decay (FID) as an example. The FID sequence contains two blocks:

[BLOCKS]
1 62 1 0 0 0 0 0
2 499947 0 0 0 0 1 0

The first block has a duration of 62 * BlockDurationRaster us and contains an RF event with an ID of 1. The block has no gradients or ADC events, indicated by zero IDs and also no extension is specified. The RF event is declared by a line:

[RF]
1 500 1 2 3 100 0 0

The RF event has a peak amplitude of 500 Hz and a delay of 100 us before starting the RF pulse. The magnitude, phase, and time compressed shapes are defined with the shape IDs of 1, 2, and 3, respectively. The RF pulse has no frequency or phase offsets.

The second block has a duration of 499947 * BlockDurationRaster us and contains an ADC event with an ID of 1, declared by:

[ADC]
1 8192 31250 29730 0 0

The ADC event contains an ADC directive with 8192 samples, a dwell time of 31250 ns, and a delay of 29730 us between the start of the block and first sample, and no frequency and phase offsets.

For more detailed specification of blocks and events, please refer to the Pulseq specification documentation at https://pulseq.github.io/specification.pdf.

3. Tutorials Overview

This section briefly introduces the tutorials one-by-one in a bit more detail.

From_FID_to_PRESS

This tutorial introduces a step-by-step design of a PRESS sequence starting from a very basic FID sequence. See 01_from_FID_to_PRESS.

Basic_gradient_echo

This tutorial demonstrates how a very basic gradient-echo sequence (GRE) can be topped up with various degrees of sophistication to achieve a desired steady-state magnetisation evolution with T1 contrast. See 02_basic_gradient_echo.

From_GRE_to_EPI

This tutorial demonstrates a possible evolution of a gradient-echo sequence (GRE) to a multi-echo GRE, segmented GRE and finally to an echo-planar sequence (EPI). See 11_from_GRE_to_EPI.

Radial_and_nonCartesian

This tutorial presents two simple radial sequences that are derived from the corresponding Cartesian counterparts. Additionally, a fast radial GRE and a very basic 2D spiral are introduced. See 12_Radial_and_nonCartesian.